+8615966019173
Black Body Infrared Calibration Furnace
Home - Products - Black Body Infrared Calibration Furnace - DY-HT2MF-300 fast cooling large surface source black body furnace, square radiation surface 300MM (room temperature +5℃-250℃)
As a standard radiation source, standard black bodies are mainly used to calibrate radiation thermometers, infrared thermometers and radiation temperature sensor detectors. Our company currently has advanced black body technology, a full range of black body products with a wide temperature range. The products are small in size, light in weight and easy to carry. They are not only suitable for calibrating radiation thermometers in laboratories of metrology institutions at all levels, but can also be used for on-site calibration.
The definition of a blackbody:an object that can absorb all external radiation energy and radiate all of its own energy.
The main technical indicators of blackbody: emissivity of blackbody, diameter of blackbody cavity, temperature uniformity and uncertainty of radiation temperature. Therefore, in order to ensure the product quality of blackbody, blackbody is usually designed according to temperature segments. Usually, the blackbody we refer to are artificial black body. The emissivity of artificial blackbody is close to 1, but not equal to 1.
The main function of a blackbody is to produce standard radiation at a certain temperature. Therefore, it is mainly used to calibrate various radiation thermometers in temperature measurement, such as optical pyrometers, infrared thermometers, infrared thermal imagers, etc.
Blackbodies are mainly used in the field of temperature measurement in industry, and the main product is black body furnace. The comparison method is usually used for the calibration and verification of radiation thermometers, that is, through a high-stability radiation source (usually a black body radiation source) and other supporting equipment, the temperature reproduced by the standard instrument is compared with the temperature reproduced by the radiation thermometer under test to determine whether it is qualified or give a calibration result.
Product Feature
1.Customized 300 large radiation surface, large effective radiation surface, uniform temperature field.
2.Human-machine interface, Chinese display, touch operation, beautiful and elegant.
3.One-key temperature setting, simple operation. Rapid heating and temperature control.
4.According to customer requirements, the language interface of multiple countries and regions can be copied for easy export.
5.It can realize fast cooling matching settings and fast cooling. (Customized optional function)
Product Advantages
1.The appearance is novel and the furnace and temperature controller are separated.
2.The heating element is insulated from the shell, and special software is used for pressure-limited heating, so no high-power power transformer is required.
3.Automatic temperature rise and temperature control is adopted, which is safe and reliable, with fast heating speed and good temperature stability.
4.It is easy to use and operate, small in size and light in weight, and easy to carry. It is not only suitable for laboratory calibration, but also for on-site calibration.
5.Large surface source isothermal blackbody radiation surface, the factory default emissivity is 0.95, and the emissivity can be adjusted between 0.96-0.99 if different emissivities are required.
6.The screen has built-in multi-point independent temperature correction function.
7.Automatic intelligent adjustment: Contains MH logic intelligent adjustment function, and advanced control algorithm for parameter self-tuning function.
8.Program control function, the equipment has 10 groups of different program setting interfaces reserved, one-key operation, convenient and fast, meeting the user's requirements for multiple detection points and improving the user's detection efficiency.
9.The communication interface can be customized RS232.
10.The device has a memory function, which allows it to quickly restore to the previous setting state after restarting, without the need for additional settings to start.
Model | DY-HT2-MF300 |
Temperature Range | RT+5℃ ~500℃ |
Operating Method | Touch Screen |
Control Mode | PID Automatic Control |
Temperature Resolution | 0.1℃ |
Radiating Surface Area | 300mm*300mm |
Emissivity | ≥0.95℃ |
Temperature Stability | ±(0.1-0.5)℃(within 10min) |
Temperature Uniformity | ±(0.2-0.8)℃(@100℃, within 2/3 area) |
Wire diameter | ≥1.5 square |
Power Supply | 220VAC 50HZ |
Power | 500W |
Communication | RS232,customizable |
Weight | Blackbody host 24kg Temperature controller 3.2kg Rapid cooling device (optional): 4.5kg |